The XHGG502A cable fault tester uses an industrial-grade 10.1-inch touchscreen all-in-one computer with a simple operating software interface, enabling human-machine interactive testing. It employs industrial-grade integrated circuits and devices, a built-in large-capacity lithium battery pack and an embedded system, providing a stable, reliable and easy-to-use fault locating solution.
This cable fault tester is a specialized instrument for measuring and analyzing the condition and fault distance of power cables. It combines modern electronic and computer technologies to achieve signal filtering, acquisition, waveform analysis and automatic fault distance calculation.
It is applicable to low-resistance, short-circuit, open-circuit and broken-circuit faults in power cables, high-frequency coaxial cables, street light cables and buried wires of various cross-sections and different media, covering low-voltage cables (6kV and below), medium and high-voltage cables (6kV-35kV) and high-voltage cables (66kV and above).
| Parameter | Specification |
|---|---|
| Sampling frequency | 1MHz ~ 400MHz |
| Low-voltage pulse amplitude | 300V ± 15% |
| Distance measurement range | ≥120km |
| Test ranges | 100m / 300m / 500m / 1km / 3km / 5km / 10km / 25km / 50km / 100km |
| Pulse width | 0.1uS / 0.30uS / 0.60uS / 1.20uS / 2.4uS / 5.0uS / 7.5uS / 10uS |
| Minimum resolution | 0.07m |
| Test blind zone | ≤10m |
| Measurement error | ≤±(0.5%×L+1m), L is the cable length |
| Pulse coupler withstand voltage | DC 35kV |
| Power supply | Rechargeable AC110V~240V, 50Hz/60Hz; built-in 11.1V lithium battery pack, charging time approximately 2 hours, working time over 5 hours |
| Operating conditions | Temperature -20°C ~ +65°C, relative humidity 90%, atmospheric pressure 750±30mmHg |
| Volume and weight | Cable fault tester: 358L×284W×168H mm, 5kg; pulse coupler: 430L×380W×220H mm, 10kg |
The XHGG502A adopts the traveling wave method testing principle. When an electromagnetic wave travels in a transmission line and encounters a point where the characteristic impedance changes, a reflected signal is generated; the polarity of the echo at the short-circuit point is opposite to that of the transmitted pulse, while the polarity of the echo at the open-circuit point is the same, allowing the operator to judge the fault nature and calculate the fault distance:
| Testing Method | Working Principle | Typical Applications |
|---|---|---|
| Traveling Wave Method (TDR) | Applies pulse signals and captures echoes at impedance-change points, converting reflection time into fault distance. | Core ranging principle covering all cable fault types |
| Low-Voltage Pulse Method | Injects 300V low-voltage pulses and analyzes the reflected waveform to measure cable length, wave speed and fault distance. | Open-circuit, low-resistance, short-circuit and broken-circuit faults |
| High-Voltage Flashover Method (Direct Flashover / Impulse Flashover) | A DC high-voltage power supply breaks down the fault point, or a high-voltage pulse signal source impulse-discharges the fault point; the sampler receives the fault reflection signal. | High-resistance flashover faults and leakage high-resistance faults |
| Multiple Pulse Method (8 times, with Pulse Coupler) | The integrated high-voltage generator and multiple pulse generator guide instantaneous high-voltage pulses to the faulty phase through the pulse coupler, displaying 8 sets of high and low voltage waveforms for comparison, effectively avoiding large cosine oscillation interference. | High-resistance faults on long-distance and high-voltage cables, especially 66kV and above transmission cables |
The instrument provides cable engineers with a comprehensive solution for power cable fault pre-location, helping minimize outage time through rapid and accurate fault distance measurement.
| Item | Quantity |
|---|---|
| Cable fault tester | 1 unit |
| Power cord | 1 piece |
| Single Q line | 1 piece |
| Double Q line | 1 piece |
| Sampling box | 1 piece |
| Coupler host | 1 unit |
| Transparent grounding wire | 1 piece |
| Coupler high voltage line A | 1 piece |
| Coupler high voltage line B | 1 piece |
| USB communication cable | 1 piece |
Overall Rating