Automatic frequency sweep from 0.1 Hz to 0.01 Hz, built for specimens whose capacitance limits a fixed-frequency set.
The XHDP-80M is the high-capacity member of the XHDP family. Instead of three fixed frequencies it sweeps automatically from 0.1 Hz down to 0.01 Hz, which lets it hold a stable output on specimens whose capacitance would overload a fixed-frequency set. The controller uses a colour touchscreen with waveform display, and a USB port is provided for exporting stored test records.
| Rating | Load capability | Typical duty |
|---|---|---|
| 30 kV | ≤ 10 µF | Cables and motors up to 10 kV |
| 40 kV (customized) | ≤ 10 µF | Cables and motors up to 10 kV |
| 50 kV | ≤ 6 µF | Cables and motors up to 15 kV |
| 60 kV | ≤ 5 µF | Cables and motors up to 25 kV |
| 80 kV | ≤ 6 µF up to 50 kV, ≤ 4 µF above 50 kV | Cables and motors up to 35 kV |
Units rated 60 kV and below use a single booster. Above 60 kV two boosters are connected in series — 25 kg for the first stage and 45 kg for the second on the 80 kV rating — which keeps each piece light enough to handle and lets either booster be used on its own for lower-voltage work. One purchase therefore covers more than one test class without adding a second complete set.
There is no frequency to select. The instrument starts at 0.1 Hz and reduces frequency as far as 0.01 Hz when specimen capacitance demands it, so the output stays inside the set's capability without the operator judging which step applies. The practical consequence is a reporting one: the frequency actually used at each step belongs in the test record, because it is part of the result rather than an operating detail.
| Parameter | Value |
|---|---|
| Output frequency | Automatic sweep, 0.1 Hz – 0.01 Hz |
| Output voltage | 30 – 80 kV peak |
| Voltage / current resolution | 0.1 kV / 0.1 mA |
| Voltage and current accuracy | 3 % |
| Peak voltage error | ≤ 3 % |
| Waveform distortion | ≤ 3 % |
| Protection operating time | ≤ 20 ms |
| High voltage lead | 150 kV rated output lead |
| Control and data | Colour touchscreen with waveform display; USB port for exporting stored records |
| Supply | 230 V ± 5 %, 50 Hz; portable generator must be a variable-frequency type |
| Service conditions | Indoor or outdoor, −10 °C to +40 °C, humidity ≤ 85 % RH |
The instrument starts at 0.1 Hz and reduces frequency as far as 0.01 Hz when the specimen capacitance demands it, so the output stays within the set's capability without the operator selecting a frequency by hand.
6 µF up to 50 kV and 4 µF above 50 kV. The 30 kV and 40 kV ratings carry 10 µF, the 50 kV rating 6 µF and the 60 kV rating 5 µF.
Each booster stays light enough to handle - 25 kg for the first stage and 45 kg for the second on the 80 kV rating - and either one can be used alone for lower-voltage work, so one set covers more than one test class.
They can be added on a customized configuration. Dielectric loss values are not published in the standard technical table because they depend on the configuration agreed for the project.
The supply is 230 V ± 5 %, 50 Hz. A conventional generator whose speed wanders will upset the boost process, so a variable-frequency generator is required where mains power is not available.
Send us the specimen capacitance per phase and the highest test voltage, and we will confirm the booster configuration and the frequency range your test list needs.